RL01/02 Pocket
Service Guide

EK-RL012-PG-003

Due to an error in the printing of the above-mentioned manual, it has become necessary to change Page 64, Paragraph 9.6, Step B.9 to read as follows:

"Place oscilloscope probe A on POS SIG and connect probe A ground lead to ground. Observe POS SIG at E11 pin 7 on version 1 or at TP 15 on version 2 and 3 of DLM. Set the vertical gain for 1 volt per division. Set oscilloscope horizontal circuit to free run (unsynced). The horizontal sweep rate is not important."

Please make the required change.